Doppler ultrasound in the measurement of pulse wave velocity: agreement with the Complior method

Abstract

Aortic stiffness is an independent predictor factor for cardiovascular risk. Different methods for determining pulse wave velocity (PWV) are used, among which the most common are mechanical methods such as SphygmoCor or Complior, which require specific devices and are limited by technical difficulty in obtaining measurements. Doppler guided by 2D ultrasound is a good alternative to these methods. We studied 40 patients (29 male, aged 21 to 82 years) comparing the Complior method with Doppler. Agreement of both devices was high (R = 0.91, 0.84-0.95, 95% CI). The reproducibility analysis revealed no intra-nor interobserver differences. Based on these results, we conclude that Doppler ultrasound is a reliable and reproducible alternative to other established methods for the measurement of aortic PWV.

Peer Review reports

Introduction

Large arteries are not simple tube conduction structures. They moderate systolic pressure increases and maintain sufficient diastolic level to guarantee myocardial perfusion. With the identification of new diseases and risk factors, it has been seen that these arteries lose their natural elasticity leading to high systolic and low diastolic blood pressure levels, which determine high pulse pressure.

Based on these premises, arterial stiffness is now considered an increasingly important biomarker in the evaluation of cardiovascular risk and the detection of incipient vascular disease. Several studies have shown that this parameter is an independent predictor of cardiovascular mortality in the elderly, hypertensive, diabetics, and patients with chronic renal failure as well as in the general population [14]. The guidelines of the European Societies of Hypertension and Cardiology (2007-2009) have postulated arterial stiffness assessment, measurement of the carotid plaque and ankle/brachial index as markers of vascular status. Any alteration of these measurements may define a state of vasculopathy that significantly increases the evaluation of risk [5].

Among the different methods of evaluating arterial stiffness, the most widely used in the literature is aortic pulse wave (PWV), specifically in the area running from the aortic arch or common carotid artery to the common femoral artery. Typically, the pulse wave is detected by pressure transducers or arterial tonometry.

The measurement of carotid-femoral PWV (Figure 1) is made by dividing the distance (from the carotid point to the femoral point) by the so-called transit time (the time of travel of the foot of the wave over the distance). Hence, PWV = D (meters)/Dt (seconds) [6, 7].

Figure 1
figure 1

Pulse wave velocity determination. Transit time is estimated by the foot-to-foot method. The foot of the wave is defined at the end of diastole, when the steep rise of the waveform begins. The transit time is the time of travel of the foot of the wave over a known distance.

Whereas the distance is a fixed parameter, the transit time has a certain variability, depending on factors such as cardiac conduction and rhythm. Given this situation, most methods take the average of several measurements.

These methods are highly reliable but have the disadvantages of requiring specific devices and software and of sometimes being impossible to perform accurately due to the difficulty in recording good pulse waves. Furthermore, the time required for the exploration is not negligible.

These disadvantages are overcome if we take the carotid-femoral PWV measurement by ultrasound, making the assumption that real pulse wave corresponds to the flow wave of spectral Doppler. In fact, this method has been used in population-based studies such as the ABC study [813]. On this basis, we designed a comparative study to assess whether PWV measured by mechanical pressure and PWV estimated by ultrasound are similar, and reliable in the measurement of arterial stiffness.

Material and methods

The subjects studied were patients from the Cardiovascular Risk and Hypertension Unit of the Dr. Josep Trueta University Hospital. Inclusion criteria were patients with essential hypertension, diabetes mellitus or chronic kidney disease and indications to testing were the evaluation of specific cardiovascular risk. Exclusion criteria were atrial fibrillation, severe cardiac valve disease and the presence of a prosthetic aorta. Both PWVc-f by the Complior® method and Doppler ultrasound measurements were performed on all patients by two investigators. Other clinical variables such as blood pressure and heart rate were also recorded. The study was approved by the Hospital Dr. Josep Trueta Ethics Committee.

PWV measured by mechanotransducers

The Complior® System (Artech Medical, Pantin, France) was used as the method of reference to determine PWVc-f, which was automatically calculated as the average of 8-10 transit times and the distance, measured from the sternal notch to the femoral artery at the groin. The test was performed in supine position, placing sensors at the carotid and femoral pulses (at the area of maximum heart rate by palpation). The result was the average of two or three speed measurements.

PWV measured by Doppler

Although it is not possible to analyze the carotid and femoral waves simultaneously, they can be normalized separately with the electrocardiogram (ECG) (gatting). We used a pulsed Doppler ultrasound with a Linear Array (6.6 MHZ) probe, synchronized with ECG and a two-second minimum sliding window (MyLab25, Esaote, Florence, Italy). The examination began with the patient in a supine position after locating the carotid artery with B-mode at the supraclavicular level (1-2 cm of the bifurcation). We then identified the wave Doppler flow simultaneously with ECG. The process was repeated on the common femoral artery in the groin. We performed three recordings of the carotid artery and three recordings of the femoral artery in the groin. Each recording involved two or three cardiac cycles. To find the transit time (TT), we measured the time from the R wave of QRS to the foot of the waveform using digital calipers (Figure 2). Six heart rate measurements were taken and the average was calculated. To determine the velocity, we used the same distance as the Complior® system.

Figure 2
figure 2

Time measurement in femoral arthery gatting with ECG in two heart rates.

To check reproducibility two blinded observers separately measured the PWV using the Doppler images of 10 consecutive patients. One observer later repeated the measurements twice at different times.

Sample size

The available sample size (n = 40) provided power >90% at 1% significance level in order to contrast the null hypothesis that the ICC equals 0.7 and the alternative hypothesis that the ICC equals 0.9 (two-sided). StudySize 2.0 Trial software was used for sample size calculations.

Statistical analysis

Continuous variables were described with mean and standard deviation and percentages were used to describe qualitative variables. Agreement between PWV measured by Doppler and Complior as well as the intra-and inteobserver agreement of the new technique was calculated based on intraclass correlation coefficient (ICC). Bland-Altmann plots were used to determine precision and bias between methods. The analysis is based on the examination of two graphs: first, the identity plot (a scatterplot of the two measurements along with the line y = x); second, the plot of the difference between methods against the gold standard (Complior).

Results

A heterogeneous group of 47 patients aged from 21 to 82 years were studied in 2009 and 2010. Seven patients were excluded, three with atrial fibrillation, three for failure to perform the Complior technique (inability to find a correct carotid pulse wave) and one due to the presence of aortic prostheses.

The sample characteristics are detailed in Table 1. PWV Complior (9.81 ± 2.76) ranged from 4.13 m/s to 19.9 m/s, similar to the PWV Doppler (9.95 ± 3.13) from 3.96 m/s to 20.2 m/s. The mean difference between the two measures was virtually the same 0.13 (0.19). The two methods showed very good agreement (R = 0.91, 0.84-0.95; CI 95%). The analysis of reproducibility also showed very good agreement (Table 2)

Table 1 Subject characteristics (n = 40)
Table 2 Agreement between methods and intraobserver and interobsever reproductibility

The identity plot in Figure 3 showed basic agreement between the two methods since the scatterplot lined up closely to the line y = x. Moreover, the Bland-Altman plot (Figure 4) indicated that the 95% limits of agreement ranged from -2 to +2. Hence, the two methods provided similar measures as the level of disagreement did not include clinically important discrepancies.

Figure 3
figure 3

Identity plot. Comparison of Doppler and Complior PWV measurement methods.

Figure 4
figure 4

Bland and Altman plot of Complior vs. Doppler PWV with mechanical method as reference. The intervals of two standard deviations are considered as the concordance limits between the two measurements, accounting for 95% of the observed differences.

Discussion

Age, atherosclerosis and the presence of certain diseases are known to decrease arterial elasticity. As can be seen in Table 3, there are many methods to estimate arterial stiffness, which has now taken on great importance in assessing cardiovascular risk. Non-invasive methods are based on local, regional or systemic measures. Direct measurement of local stiffness is usually made by echotracking systems, which measure in situ the differences of arterial diameter over the wave flow.

Table 3 Devices and methods used to determine arterial stiffness

The most commonly used regional method is aortic pulse wave velocity, which requires two variables: the distance between two points in the artery and the time taken by the pulse to cover that distance. Carotid-femoral PWV is a simple, non-invasive, robust and reproducible method that is regarded as the gold standard for measuring arterial stiffness since epidemiological studies have found it to be an independent predictor of cardiovascular events.

The Complior System®, which uses two mechanotransducers applied to the skin and measures real-time pulse waves at carotid and femoral points, is used in most of these studies. Another widely used system is SphygmoCor®, which uses an applanation tonometer [616]. This device can also calculate the central pressure and augmentation index.

Complior records both waves simultaneously, whereas SphygmoCor records consecutively using ECG. In this case, changes in heart rate between two recordings may determine a variation in transit time.

These mechanical methods have the disadvantages that there is a prolonged learning period in order to become an experienced observer and that the devices used lack versatility. Furthermore, the technical difficulty in obtaining measurements and anatomical limitations of some patients make it necessary to find faster and more versatile methods for measuring PWV [17]. The measurement of stiffness by ultrasound has the advantages of being significantly quicker and of not requiring a specific device.

Several studies, the largest of which was undertaken by Sutton-Tyrrell et al. and included 2488 patients, have used Doppler ultrasound to measure aortic PWV. Table 4 sets out the main PWV studies together with the methods employed [18]. However, few studies have compared a mechanical method with the use of Doppler [19].

Table 4 Studies with aortic PWV [18]

This study demonstrates that Doppler ultrasound can be used to measure aortic PWV in a reliable and reproducible way, giving similar results to Complior®, which we took as a gold-standard. In addition, B-mode ultrasound provides an anatomical image that can increase the precision of measurements (for example, using the carotid or femoral bifurcation as a reference). This method has the further advantages of shorter performance time, short learning curve and the absence of anatomical limitations, which are especially pronounced in the carotid artery. The versatility of ultrasound also permits us to explore simultaneously other pathologies such as plaques or blockages in the carotid and femoral territories as well as to assess intima-media thickness.

Conclusions

The aim of this study was to consider the use of Doppler as an alternative to a more established method (Complior®) of measuring pulse wave velocity and to highlight certain advantages that Doppler has over this reference technique. Although the number of patients studied in this first study to specifically compare these two devices is limited, the findings are sufficiently powerful to demonstrate a high correlation between the two systems and so to justify the use of Doppler ultrasound both in clinical practice and clinical studies to assess arterial stiffness. However, automated methods should be developed for calculating transit time to reduce the variability from the use of manual calipers.

References

  1. Safar ME, London GM, Plante GE: Arterial stiffness and kidney function. Hypertension. 2004, 43 (2): 163-68. 10.1161/01.HYP.0000114571.75762.b0

    Article  CAS  PubMed  Google Scholar 

  2. Laurent S, Boutouyrie P: Arterial stiffness: a new surrogate end point for cardiovascular disease?. J Nephrol. 2007, 20 Suppl 12:S45-50.

    Google Scholar 

  3. DeLoach SS, Townsend RR: Vascular stiffness: its measurement and significance for epidemiologic and outcome studies. Clin J Am Soc Nephrol. 2008, 3 (1): 184-192. 10.2215/CJN.03340807

    Article  PubMed  Google Scholar 

  4. Gosse Philippe; Roche Frederic; Dauphinot Virginie; Maudoux Delphine; Pichot Vincent; Barthelemy Jean Claude: Components of arterial stiffness in a population of 65-year-old subjects: PROOF study. Journal of Hypertension. 2008, 26 (6): 1138-46. 10.1097/HJH.0b013e3282fa142b

    Article  PubMed  Google Scholar 

  5. 2007 Guidelines for the management of arterial hypertension. Eur Heart J. 2007, 28: 1462-1536.

  6. Laurent S, Cockcroft J, Van Bortel L, Boutouyrie P, Giannattasio C, Hayoz D, Pannier B, Vlachopoulos C, Wilkinson I, Struijker-Boudier H: European Network for Non-invasive Investigation of Large Arteries. Expert consensus document on arterial stiffness: methodological issues and clinical applications. Eur Heart J. 2006, 27 (21): 2588-605. 10.1093/eurheartj/ehl254

    Article  PubMed  Google Scholar 

  7. Weber Thomas, Marcus Ammer, Martin Rammer, Audrey Adji, O'Rourke Michael F, Wassertheurer Siegfried, Rosenkranz Stefan, Eber Bernd: Noninvasive determination of carotid-femoral pulse wave velocity depends critically on assessment of travel distance: a comparison with invasive measurement. Journal of Hypertension. 2009, 27 (8): 1624-30. 10.1097/HJH.0b013e32832cb04e

    Article  CAS  PubMed  Google Scholar 

  8. Vaitkevicius PV, Fleg JL, Engel JH, O'Connor FC, Wright JG, Lakatta LE, Yin FC, Lakatta EG: Effects of age and aerobic capacity on arterial stiffness in healthy adults. Circulation. 1993, 88 (4 Pt 1): 1456-62.

    Article  CAS  PubMed  Google Scholar 

  9. Sutton-Tyrrell K, Mackey RH, Holubkov R, Vaitkevicius PV, Spurgeon HA, Lakatta EG: Measurement variation of aortic pulse wave velocity in the elderly. Am J Hypertens. 2001, 14 (5 Pt 1): 463-8. 10.1016/S0895-7061(00)01289-9

    Article  CAS  PubMed  Google Scholar 

  10. Cruickshank K, Riste L, Anderson SG, Wright JS, Dunn G, Gosling RG: Aortic pulse-wave velocity and its relationship to mortality in diabetes and glucose intolerance: an integrated index of vascular function?. Circulation. 2002, 106 (16): 2085-90. 10.1161/01.CIR.0000033824.02722.F7

    Article  PubMed  Google Scholar 

  11. Sutton-Tyrrell K, Najjar SS, Boudreau RM, Venkitachalam L, Kupelian V, Simonsick EM, Havlik R, Lakatta EG, Spurgeon H, Kritchevsky S, Pahor M, Bauer D, Newman A, Health ABC Study: Elevated Aortic Pulse Wave Velocity, a Marker of Arterial Stiffness, Predicts Cardiovascular Events in Well-Functioning Older Adults. Circulation. 2005, 111: 3384-90. 10.1161/CIRCULATIONAHA.104.483628

    Article  PubMed  Google Scholar 

  12. Rajzer MW, Wojciechowska W, Klocek M, Palka I, Brzozowska-Kiszka M, Kawecka-Jaszcz K: Comparison of aortic pulse wave velocity measured by three techniques: Complior, SphygmoCor and Arteriograph. J Hypertens. 2008, 26 (10): 2001-7. 10.1097/HJH.0b013e32830a4a25

    Article  CAS  PubMed  Google Scholar 

  13. Baguet JP, Kingwell BA, Dart AL, Shaw J, Ferrier KE, Jennings GL: Analysis of the regional pulse wave velocity by Doppler: methodology and reproducibility. J Hum Hypertens. 2003, 17 (6): 407-412. 10.1038/sj.jhh.1001566

    Article  PubMed  Google Scholar 

  14. Pannier BM, Avolio AP, Hoeks A, Mancia G, Takazawa K: Methods and devices for measuring arterial compliance in humans. Am J Hypertens. 2002, 15 (8): 743-753. 10.1016/S0895-7061(02)02962-X

    Article  PubMed  Google Scholar 

  15. Huck CJ, Bronas UG, Williamson EB, Draheim CC, Duprez DA, Dengel DR: Noninvasive measurements of arterial stiffness: repeatability and interrelationships with endothelial function and arterial morphology measures. Vasc Health Risk Manag. 2007, 3 (3): 343-349.

    PubMed  PubMed Central  Google Scholar 

  16. Ibrahim el-SH, Johnson KR, Miller AB, Shaffer JM, White RD: Measuring aortic pulse wave velocity using high-field cardiovascular magnetic resonance: comparison of techniques. J Cardiovasc Magn Reson. 2010, 12 (1): 26- 10.1186/1532-429X-12-26

    Article  PubMed Central  Google Scholar 

  17. Matthys Koen, Pascal Verdonck: Development and modelling of arterial applanation tonometry: A review. Technol Health Care. 2002, 10 (1): 65-76.

    PubMed  Google Scholar 

  18. Vlachopoulos C, Aznaouridis K, Stefanadis C: Prediction of cardiovascular events and all-cause mortality with arterial stiffness: a systematic review and meta-analysis. J Am Coll Cardiol. 2010, 55 (13): 1318-1327. 10.1016/j.jacc.2009.10.061

    Article  PubMed  Google Scholar 

  19. Jiang B, Liu B, McNeill KL, Chowienczyk PJ: Measurement of pulse wave velocity using pulse wave Doppler ultrasound: comparison with arterial tonometry. Ultrasound Med Biol. 2008, 34 (3): 509-512. 10.1016/j.ultrasmedbio.2007.09.008

    Article  PubMed  Google Scholar 

  20. Anderson SG, Sanders TA, Cruickshank JK: Plasma fatty acid composition as a predictor of arterial stiffness and mortality. Hypertension. 2009, 53: 839-45. 10.1161/HYPERTENSIONAHA.108.123885

    Article  CAS  PubMed  Google Scholar 

  21. Blacher J, Guerin AP, Pannier B, Marchais SJ, Safar ME, London GM: Impact of aortic stiffness on survival in end-stage renal disease. Circulation. 1999, 99: 2434-9.

    Article  CAS  PubMed  Google Scholar 

  22. Boutouyrie P, Tropeano AI, Asmar R, Gautier I, Benetos A, Lacolley P, Laurent S: Aortic stiffness is an independent predictor of primary coronary events in hypertensive patients. A longitudinal study. Hypertension. 2002, 9: 10-5. 10.1161/hy0102.099031.

    Article  Google Scholar 

  23. Choi CU, Park EB, Suh SY, Kim JW, Kim EJ, Rha SW, Seo HS, Oh DJ, Park CG: Impact of aortic stiffness on cardiovascular disease in patients with chest pain. Assessment with direct intra-arterial measurement. Am J Hypertens. 2007, 20: 1163-9.

    PubMed  Google Scholar 

  24. Laurent S, Boutouyrie P, Asmar R, Gautier I, Laloux B, Guize L, Ducimetiere P, Benetos A: Aortic stiffness is an independent predictor of all-cause and cardiovascular mortality in hypertensive patients. Hypertension. 2001, 37: 1236-41.

    Article  CAS  PubMed  Google Scholar 

  25. Mattace-Raso FU, van der Cammen TJ, Hofman A, van Popele NM, Bos ML, Schalekamp MA, Asmar R, Reneman RS, Hoeks AP, Breteler MM, Witteman JC: Arterial stiffness and risk of coronary heart disease and stroke. The Rotterdam Study. Circulation. 2006, 113: 657-63. 10.1161/CIRCULATIONAHA.105.555235

    Article  PubMed  Google Scholar 

  26. Meaume S, Benetos A, Henry OF, Rudnichi A, Safar ME: Aortic pulse wave velocity predicts cardiovascular mortality in subjects >70 years of age. Arterioscler Thromb Vasc Biol. 2001, 21: 2046-50. 10.1161/hq1201.100226

    Article  CAS  PubMed  Google Scholar 

  27. Mitchell GF, Hwang SJ, Vasan RS, Larson MG, Pencina MJ, Hamburg NM, Vita JA, Levy D, Benjamin EJ: Arterial stiffness and cardiovascular events. The Framingham Heart Study. Circulation. 2010, 121: 505-11. 10.1161/CIRCULATIONAHA.109.886655

    Article  PubMed  PubMed Central  Google Scholar 

  28. Pannier B, Guerin AP, Marchais SJ, Safar ME, London GM: Stiffness of capacitive and conduit arteries: prognostic significance for end-stage renal disease patients. Hypertension. 2005, 45: 592-6. 10.1161/01.HYP.0000159190.71253.c3

    Article  CAS  PubMed  Google Scholar 

  29. Shoji T, Emoto M, Shinohara N, Kakiya R, Tsujimoto Y, Kishimoto H, Ishimura E, Tabata T, Nishizawa Y: Diabetes mellitus, aortic stiffness, and cardiovascular mortality in end-stage renal disease. J Am Soc Nephrol. 2001, 12: 2117-24.

    CAS  PubMed  Google Scholar 

  30. Shokawa T, Imazu M, Yamamoto H, Toyofuku M, Tasaki N, Okimoto T, Yamane K, Kohno N: Pulse wave velocity predicts cardiovascular mortality. Findings from the Hawaii-Los Angeles-Hiroshima Study. Circ J. 2005, 69: 259-64. 10.1253/circj.69.259

    Article  PubMed  Google Scholar 

  31. Terai M, Ohishi M, Ito N, Takagi T, Tatara Y, Kaibe M, Komai N, Rakugi H, Ogihara T: Comparison of arterial functional evaluations as a predictor of cardiovascular events in hypertensive patients: the Non-Invasive Atherosclerotic Evaluation in Hypertension (NOAH) Study. Hypertens Res. 2008, 31: 1135-45. 10.1291/hypres.31.1135

    Article  PubMed  Google Scholar 

  32. Wang KL, Cheng HM, Sung SH, Chuang SY, Li CH, Spurgeon HA, Ting CT, Najjar SS, Lakatta EG, Yin FC, Chou P, Chen CH: Wave reflection and arterial stiffness in the prediction of 15-year all-cause and cardiovascular mortalities: a community-based study. Hypertension. 2010, 55: 799-805. 10.1161/HYPERTENSIONAHA.109.139964

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  33. Willum-Hansen T, Staessen JA, Torp-Pedersen C, Rasmussen S, Thijs L, Ibsen H, Jeppesen J: Prognostic value of aortic pulse wave velocity as index of arterial stiffness in the general population. Circulation. 2006, 113: 664-70. 10.1161/CIRCULATIONAHA.105.579342

    Article  PubMed  Google Scholar 

  34. Zoungas S, Cameron JD, Kerr PG, Wolfe R, Muske C, McNeil JJ, McGrath BP: Association of carotid intima-medial thickness and indices of arterial stiffness with cardiovascular disease outcomes in CKD. Am J Kidney Dis. 2007, 50: 622-30. 10.1053/j.ajkd.2007.07.012

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

We thank Angie Lopez and Mirta Sola, for outstanding nursing and technical support.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jordi Calabia.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

JC designed the study, supervised subject recruitment, carried out collection of data and part of assessment of PWV by doppler and wrote the manuscript. PT carried out assessment of PWV by Complior, assisted in recruitment and manuscript revision. MG helped in the interpretation of the results and statistical analysis. IG assisted data interpretation and wrote the manuscript. NM and DF assisted recruitment, data collection and manuscript revision. BG assisted in assessment of PWV by both methods and participated in the analysis of reproducibility. MV assisted in study design, data interpretation and manuscript revision. All authors read and approved the manuscript.

Authors’ original submitted files for images

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Calabia, J., Torguet, P., Garcia, M. et al. Doppler ultrasound in the measurement of pulse wave velocity: agreement with the Complior method. Cardiovasc Ultrasound 9, 13 (2011). https://doi.org/10.1186/1476-7120-9-13

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1476-7120-9-13

Keywords