Variability in echocardiographic measurements of left ventricular function in septic shock patients

Abstract

Background

Echocardiography is increasingly used for haemodynamic evaluation and titration of therapy in intensive care, warranting reliable and reproducible measurements. The aim of this study was to evaluate the observer dependence of echocardiographic findings of left ventricular (LV) diastolic and systolic dysfunction in patients with septic shock.

Methods

Echocardiograms performed in 47 adult patients admitted with septic shock to a general intensive care unit (ICU) were independently evaluated by one cardiologist and one intensivist for the following signs: decreased diastolic tissue velocity of the base of the LV septum (é), increased early mitral inflow (E) to é ratio (E/é), decreased LV ejection fraction (EF) and decreased LV global longitudinal peak strain (GLPS). Diastolic dysfunction was defined as é <8.0 cm/s and/or E/é ≥15 and systolic dysfunction as EF <50% and/or GLPS > −15%. Ten randomly selected examinations were re-analysed two months later. Pearson’s r was used to test the correlation and Bland-Altman plots to assess the agreement between observers. Kappa statistics were used to test the consistency between readers and intraclass correlation coefficients (ICC) for inter- and intraobserver variability.

Results

In 44 patients (94%), image quality was sufficient for echocardiographic measurements. The agreement between observers was moderate (k = 0.60 for é, k = 0.50 for E/é and k = 0.60 for EF) to good (k = 0.71 for GLPS). Pearson’s r was 0.76 for é, 0.85 for E/é, 0.78 for EF and 0.84 for GLPS (p < 0.001 for all four). The ICC between observers for é was very good (0.85; 95% confidence interval (CI) 0.73-0.92), good for E/é (0.70; 95% CI 0.45 – 0.84), very good for EF (0.87; 95% CI 0.77 – 0.93), excellent for GLPS (0.91; 95% CI 0.74 – 0.95), and very good for all measures repeated by one of the observers. On Bland-Altman analysis, the mean differences and 95% limits of agreement for é, E/é, EF and GLPS were −0.01 (0.04 – 0.07), 2.0 (−14.2 – 18.1), 0.86 (−16 – 14.3) and 0.04 (−5.04 – 5.12), respectively.

Conclusions

Moderate observer-related differences in assessing LV dysfunction were seen. GLPS is the least user dependent and most reproducible echocardiographic measurement of LV function in septic shock.

Peer Review reports

Background

Cardiac dysfunction is a well-known complication of septic shock. It was first described using radionuclide cineangiography [1], and has been studied further since the introduction of echocardiography in intensive care clinical practice. Numerous studies have described diastolic as well as systolic dysfunction, or a combination of the two, in septic shock, using a variety of different echocardiographic parameters [2-6]. Furthermore, echocardiography is now increasingly used for haemodynamic monitoring and in titrating therapy in septic shock [7].

The frequency of cardiac dysfunction in septic shock as well as its impact on mortality has differed vastly between studies [2,3,5]. Differences in study size, inclusion criteria and in the prevalence of pre-existing heart-disease may have contributed to the varying results. In cardiology settings, a substantial interobserver variability in echocardiographic assessment has been described, especially in stress echocardiography [8-10]. In intensive care, echocardiographic assessment is often complicated by tachycardia, high levels of catecholamines and difficulties in image acquisition, not unlike that of stress echocardiography. However, the addition of observer dependency to the interpretation of the echocardiographic findings, and its impact on the differing results reported, has been scarcely studied [9,11,12].

In this study, we assess the reliability of echocardiographic parameters used to indicate left ventricular dysfunction in patients with septic shock. Our aim is to identify the least observer-dependent echocardiographic parameter describing diastolic and systolic left ventricular function.

Methods

This study was approved by the Regional Ethical Review Board in Linköping, Sweden (Dnr 2012/233-31). When possible, informed consent was sought from patients at inclusion. Due to the observational nature of the study we were permitted to assume consent in patients who because of acute illness were unable to give their permission. In these cases, informed consent was obtained as soon as possible after recovery.

The dataset used for this study was originally collected for a study assessing speckle tracking echocardiography in septic shock patients [13]. Patients aged 18 years or older, admitted to the mixed non-cardiothoracic ICU of Linköping University Hospital, presenting with septic shock and with an expected ICU stay of 24 hours or longer, were screened for eligibility. In total, 50 patients were included from October 2012 to September 2014. Patients could be included only once. Patients who by the treating physician were not expected to survive longer than 24 hours, in whom intensive care treatment was partly withheld from admission, or who due to language barriers or mental inability were not expected to be able to give consent even after recovery, were excluded.

Septic shock was defined according to the Surviving Sepsis Campaign (SSC) criteria, and all patients were managed in compliance with the SSC guidelines for the treatment of septic shock [14]. Patients were considered to have a history of cardiac disease if they had prior or current ischaemic heart disease, cardiac surgery, hypertension or cardiac failure.

Transthoracic echocardiography was performed on the day of admission, all by an expert echocardiographer not involved in patient care (JE). A Vivid E9 ultrasound scanner was used, acquiring two-dimensional (2D) apical two-chamber, four-chamber and long axis views (2C, 4C and ALAX) of the left ventricle at a frame rate of >40 frames/sec. Images were analysed offline using dedicated software (EchoPac BT 112, GE Ultrasound, Horten, Norway), independently by two observers (LDG and JE). E-velocity was measured using pulsed wave (PW) Doppler in the mitral inflow at the tip of the valve. Diastolic tissue velocity of the base of the LV septum (é) was measured in the apical 4C view using PW tissue Doppler, and E/é ratios were calculated. Left ventricular (LV) volumes and ejection fraction (EF) were calculated using the modified biplanar Simpson’s method. Global longitudinal peak strain (GLPS) was calculated as the average speckle tracking strain from the 18 LV segments from the 2C, 4C and ALAX views (six segments per view in three different views). All echocardiographic studies were recorded over three consecutive cardiac cycles, independently of breathing cycles, and averaged. In patients with non-sinus rhythm measurements were collected and averaged over 5 – 10 heartbeats. Diastolic dysfunction was defined as E/é >15 and/or é <8 cm/s [15]. Systolic dysfunction was defined as EF <50% [16], and GLPS was considered decreased when > −15% [17]. A random sample of 10 examinations were re-analysed two months later by one of the observers (LDG) in order to study intra-observer repeatability.

Clinical data on co-morbidities and on ICU length of stay and outcome were all collected prospectively. To assess severity of illness, Simplified Acute Physiology Score 3 (SAPS 3) and Sequential Organ Failure Assessment (SOFA) score were calculated on admission.

Statistical analysis

Data are presented as medians (lower quartile – upper quartile), numbers (percentages), and means with standard deviations (range), as appropriate. Inter-observer variability of echocardiographic parameters was determined by the intra-class correlation coefficient, Pearson’s correlation coefficient and Bland-Altman plots [18]. Reliability analyses using kappa statistics were performed to determine consistency between observers regarding the presence or absence of diastolic or systolic dysfunction. The kappa value for agreement was interpreted as follows: poor <0.20, fair, 0.21 – 0.40; moderate, 0.41 – 0.60; good, 0.61 – 0.80; and very good, 0–81 – 1.0 [19]. Intra-observer repeatability was calculated using intra-class correlation coefficient. All probability values are two-tailed and significance was set at p < 0.05. All statistical analyses were performed using IBM SPSS v22.0 (IBM Corp, Armonk, NY, USA).

Results

Fifty patients in septic shock were included. Two patients died before echocardiography could be undertaken and in one patient, images were lost in the storage process. These three patients were excluded from the statistical analysis. In the remaining 47 patients, echocardiograms from the day of ICU admission were available. Twenty-three of the patients (49%) had previously known cardiac comorbidities, but there were no patients with severe valvular regurgitation, and none with endocarditis. A three-lead electrocardiogram (ECG) recorded for monitoring purposes was analysed regarding cardiac rhythm, but 12-lead ECG was not performed. Table 1 summarizes the main clinical characteristics of studied patients.

Table 1 Baseline and echocardiographic characteristics of studied patients

In three patients (6%), image quality was too poor to allow echocardiographic analysis. In the remaining 44 patients (94%), E, é, LVEF and GLPS could be assessed. Table 2 shows the diastolic echocardiographic parameters obtained by the two observers. Of the 18 LV segments, GLPS could be measured in, on average, 17 segments by both observers (range 13 – 18 vs range 12 – 18).

Table 2 Left ventricular function on day 1

The é measured by the two observers was 8.1 ± 3.0 and 9.0 ± 4.1, respectively (Table 2). On Pearson’s correlation test, r = 0.76 (p < 0.001, Figure 1), and the intraclass correlation was 0.85 (95% CI 0.20 – 0.73, Table 3). On Bland-Altman analysis of é measurements, the mean difference was 1.4 cm/s and the standard deviation (SD) 2.7 (Figure 1). Diastolic dysfunction defined as é <8.0 cm/s was identified in 21 (45%) and 18 (38%) patients, respectively. The observers agreed on diastolic dysfunction being present 15 patients (71% of those where at least one observer had indicated the presence of diastolic dysfunction). The inter-observer agreement was found to be kappa 0.60 (95% CI 0.36 – 0.83).

Figure 1
figure 1

The correlation between and Bland-Altman analysis of é and E/é measured by two observers. Pearson’s correlation coefficient, mean measurements, differences and 95% limits of agreement (1.96 SD) between measurements by two observers.

Table 3 Repeatability of results

The E/é ratio derived from measurements by the two observers was 14.4 ± 11.7 and 11.4 ± 8.9 (Table 1), r was 0.85 (p < 0.001, Figure 1) and the intraclass correlation coefficient 0.71 (95% CI 0.46 – 0.84, Table 2). The Bland-Altman plot in Figure 1 shows the mean difference being 2.1 (SD 8.25). When LV diastolic dysfunction was defined as E/é >15, the observers identified LV diastolic dysfunction in 11 (23%) and 6 (13%) patients, respectively, with agreement in 5 (45%), and the kappa was 0.50 (95% CI 0.19 – 0.80).

EF measured by the two observers was 49 ± 11 and 50 ± 12, r = 0.78 (p < 0.001, Figure 2), the intraclass correlation coefficient was 0.87 (95% CI 0.77 – 0.93, Table 2) and the mean difference was 0.9% (SD 7.7, Figure 2). Decreased EF, defined as <50%, was considered present in 24 (51%) vs 21 (45%) patients. The observers agreed on EF being decreased in 18 patients (75%), yielding a kappa of 0.60 (95% CI 0.36 – 0.83).

Figure 2
figure 2

The correlation between and Bland-Altman analysis of EF and GLPS measured by two observers. Pearson’s correlation coefficient, mean measurements, differences and 95% limits of agreement (1.96 SD) between measurements by two observers.

The mean GLPS was −16.6 ± 4.4 and −16.6 ± 4.6, when measured by the two observers. The correlation coefficient (r) was 0.84 (p < 0.001, Figure 2) and the intraclass correlation coefficient was 0.91 (95% CI 0.84 – 0.95). An example of observed inter- and intraobserver reproducibility is illustrated in Figure 3 and is further highlighted in Additional file 1. The Bland-Altman plot in Figure 2 shows a mean difference of 0.04% (SD 2.6). The observers found GLPS decreased in 17 (36%) and 15 (32%) patients, respectively, agreeing in 13 (76%), and the inter-rater reliability was found to be kappa 0.71 (95% CI 0.49 – 0.92). Table 3 summarizes the repeatability of results, showing the intraclass correlation coefficients of inter- and intraobserver variability of LV function parameters studied.

Figure 3
figure 3

Variability of GLPS measurements. Segmental strain and GLPS of the left ventricle illustrated using automatic function imaging from three apical views and averaged for 18 segments; A. Measurements performed by observer 1, GLPS −17.9%; B. Measurements performed by observer 2, GLPS −20.0%; and C. Measurements re-assessed by observer 2, GLPS −18.4%.

Thirteen patients had atrial fibrillation when the echocardiographic examination was performed. When these were excluded from the analysis, Pearson’s correlation for é, E/é, EF and GLPS was 0.72, 0.80, 0.79 and 0.81, respectively (in all cases p < 0.001). ICC (95% CI) improved slightly to 0.83 (0.64 – 0.92), 0.88 (0.75 – 0.95), 0.88 (0.76 – 0.94) and 0.90 (0.79 – 0.95) for é, E/é, EF and GLPS. On Bland-Altman analyses, the mean differences and 95% limits of agreement (1.96 SD) for é, E/é, EF and GLPS were −1.8 (−7.8 – 4.2), 3.82 (−6.8 –14.5), −0.26 (−15.2 – 14.7) and 0.09 (−5.5 – 5.7), respectively.

Discussion

In this study, we aimed at analysing the observer dependence of echocardiographic signs of cardiac dysfunction in patients with septic shock to establish the most robust, user independent parameter. Our findings suggest that there are moderate inter-individual differences regarding the measurement of diastolic and systolic parameters used to identify cardiac dysfunction in septic shock. GLPS is the least user-dependent and most reproducible parameter.

LV dysfunction in septic shock has been described with varying incidence. Impaired diastolic function with decreased é or increased E/é, as well as systolic dysfunction with decreased EF, has been reported in 20 – 60% of patients. A direct comparison between studies is complicated by differences in study size and in the prevalence of pre-existing cardiac disease, but also by the lack of true consensus on the definition of LV dysfunction in septic shock. In cardiology settings, the impact of observer dependence on echocardiographic measurements regarding diastolic as well as systolic parameters has been studied [8-10], but remarkably less so in the intensive care setting. Thus, to what extent the differing results regarding prevalence and implications on outcome in septic shock depends on the reading of the echocardiograms is unknown.

The echocardiographic measures to assess diastolic myocardial function, in this study é and E/é, reflect different aspects of LV filling, each with potential limitations. Furthermore, the measurements are manually obtained and therefore inherently subjective. Also, the accuracy and reproducibility of é and E/é is debated in patients with atrial fibrillation (AF), ischaemic heart disease or hypertrophic cardiomyopathy [20-22]. The prevalence of AF and other heart diseases in this study is in the same range as in other studies on septic shock patients [2,23]. In patients with AF, measurements were performed and averaged over a larger number of cardiac cycles. Still, our results indicate that atrial fibrillation hampers the consistency between measurements, possibly explaining some of the observed differences in previous studies.

The accuracy of é as well as E/é has been shown to be dependent on changes in volume loading and ventilator settings [24-26], possibly yielding different results even within the same echocardiographic examination and further highlighting the difficulties associated with the clinical intensive care situation. In heart failure patients, the observer concordance of diastolic measurements has been shown to be less than ideal [8]. Our results are in the same range as in this previous study, possibly reflecting similar difficulties in assessing diastolic function in septic shock patients as in those with heart failure. Interestingly, diastolic dysfunction in septic shock has, albeit with conflicting results, been shown to increase mortality [5,23]. Observer-related variability, such as shown here, may have contributed to the vastly differing results reported regarding diastolic function and its impact on mortality in septic shock patients.

For evaluating systolic LV function, we used the biplane Simpson method which is widely recommended and has been shown to reliably detect even small changes in EF in ischaemic heart disease [16,27]. One previous study has focused on the repeatability of LV systolic function parameters in intensive care patients in shock, the majority of which were in septic shock [11]. In this study inter- and intraobserver reproducibility was within a range that was considered acceptable, but the use of Simpson’s single plane method in addition to other measures of LV systolic function complicates direct comparison with our study. We found the agreement between observers to be moderate when defining EF as decreased if below 50%, in contrast to the very good correlation between observers regarding absolute values. This may illustrate the effect of an arbitrary cut-off, and may also be a reflection of the varying incidence and the unclear impact on mortality of decreased EF in septic shock, recently reported in a meta-analysis [28]. Interestingly, the presence of atrial fibrillation had little impact on the concordance of the results regarding systolic measurements.

In stress echocardiography a high degree of variability has been demonstrated even among expert observers [9]. A recent study has shown a much improved reproducibility of results in the interpretation of dobutamine stress echocardiography when using GLPS [10]. In line with this, our findings indicate a good reproducibility of GLPS in septic shock, with a small, scarcely clinically relevant, difference between observers and without obvious bias. There are indeed similarities between the context of stress echocardiography and that of echocardiography in septic shock patients, both situations being challenged by difficulties in image acquisition, high levels of endogenous and exogenous catecholamines in addition to tachycardia and arrhythmias. The superior reproducibility of GLPS in stress echocardiography, also shown here in septic shock, may reflect the previously described advantage of a semi-automated technique with less manual processing [29]. GLPS may thus represent an objective and possibly less user-dependent method than other echocardiographic measurements.

Our findings have clinical implications since echocardiography is used for the diagnosis of cardiac dysfunction in septic shock. Furthermore, echocardiography is increasingly recommended for circulatory monitoring in septic shock patients [7,30]. Since we found considerable interobserver variability in diastolic measurements, our results imply the need for some caution in using them to guide therapy in this group of patients. Much of the treatment recommended in septic shock predominantly targets the systolic function of the heart [14]. The good repeatability of systolic function parameters found here may further encourage the use of echocardiography as a monitoring tool. However, our results suggest that echocardiographic findings should be interpreted with some caution, and in relation to other laboratory and clinical data in the individual patient. Furthermore, GLPS has recently been shown to be a sensitive marker of LV dysfunction in septic shock [6,13,31]. This fact, in addition to the strong reliability and reproducibility of GLPS shown here, indicates a future clinical role of GLPS as a marker of LV function in septic shock.

Limitations

The selection of patients has a heavy influence on the results of echocardiographic assessment. In this study, patients expected to survive less than 24 hours were excluded, and this may have contributed to the high percentage of successful acquisitions. LV function parameters used in this study are all well established in patients with cardiac disease, but there is no clear definition on cardiac dysfunction in septic shock. Neither is there any consensus on what degree of variability in measurements is acceptable in this group of patients, since this aspect of echocardiographic assessment has not been addressed before. Furthermore, the number of patients studied is limited and the dataset used was not originally collected for these analyses.

A strength of the study is the fact that images were collected by an expert-level echocardiographer. Thus, the focus of the study is on the repeatability of measurements, irrespective of image acquisition. Moreover, all echocardiographic analyses were performed independently and while blinded to patient identity and clinical background, and the samples used for comparison are therefor of equal size. Further strength lies in the systematic approach to assessing inter- as well as intraobserver reliability and reproducibility.

Conclusion

We conclude that echocardiographic measurement of LV function is feasible and reproducible in the majority of septic shock patients. Because of the observer-related differences in the measurements used, results should be interpreted in relation to other clinical findings when used for guiding therapy. GLPS is the least user dependent and most reproducible echocardiographic finding of LV function in septic shock. GLPS may thus represent a useful tool in the evaluation of LV function in this group of patients.

Abbreviations

2D:

Two-dimensional

2C:

Two-chamber view

4C:

Four-chamber view

AF:

Atrial fibrillation

ALAX:

Apical long-axis view

CI:

Confidence interval

é:

Early mitral tissue Doppler velocity

E/é:

Early mitral inflow to early mitral tissue Doppler velocity

ECG:

Electrocardiogram

EF:

Ejection fraction

GLPS:

Global longitudinal peak strain

ICU:

Intensive care unit

IQR:

Inter-quartile range

LV:

Left ventricular

PW:

Pulsed wave Doppler

SAPS3:

Simplified acute physiology score 3

SD:

Standard deviation

SOFA:

Sequential organ failure assessment

SSC:

Surviving Sepsis Campaign

References

  1. Parker MM, Shelhamer JH, Bacharach SL, Green MV, Natanson C, Frederick TM, et al. Profound but reversible myocardial depression in patients with septic shock. Ann Intern Med. 1984;100:483–90.

    Article  CAS  PubMed  Google Scholar 

  2. Pulido JN, Afessa B, Masaki M, Yuasa T, Gillespie S, Herasevich V, et al. Clinical spectrum, frequency, and significance of myocardial dysfunction in severe sepsis and septic shock. Mayo Clin Proc. 2012;87:620–8.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Weng L, Liu YT, Du B, Zhou JF, Guo XX, Peng JM, et al. The prognostic value of left ventricular systolic function measured by tissue Doppler imaging in septic shock. Crit Care. 2012;16:R71.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Vieillard-Baron A, Caille V, Charron C, Belliard G, Page B, Jardin F. Actual incidence of global left ventricular hypokinesia in adult septic shock. Crit Care Med. 2008;36:1701–6.

    Article  PubMed  Google Scholar 

  5. Landesberg G, Gilon D, Meroz Y, Georgieva M, Levin PD, Goodman S, et al. Diastolic dysfunction and mortality in severe sepsis and septic shock. Eur Heart J. 2012;33:895–903.

    Article  CAS  PubMed  Google Scholar 

  6. Orde SR, Pulido JN, Masaki M, Gillespie S, Spoon JN, Kane GC, et al. Outcome prediction in sepsis: speckle tracking echocardiography based assessment of myocardial function. Crit Care. 2014;18:R149.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Vincent JL, Rhodes A, Perel A, Martin GS, Della Rocca G, Vallet B, et al. Clinical review: update on hemodynamic monitoring–a consensus of 16. Crit Care. 2011;15:229.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Unzek S, Popovic ZB, Marwick TH, Diastolic Guidelines Concordance I. Effect of recommendations on interobserver consistency of diastolic function evaluation. JACC Cardiovasc Imaging. 2011;4:460–7.

    Article  PubMed  Google Scholar 

  9. Hoffmann R, Lethen H, Marwick T, Arnese M, Fioretti P, Pingitore A, et al. Analysis of interinstitutional observer agreement in interpretation of dobutamine stress echocardiograms. J Am Coll Cardiol. 1996;27:330–6.

    Article  CAS  PubMed  Google Scholar 

  10. Yamada A, Luis SA, Sathianathan D, Khandheria BK, Cafaro J, Hamilton-Craig CR, et al. Reproducibility of regional and global longitudinal strains derived from two-dimensional speckle-tracking and doppler tissue imaging between expert and novice readers during quantitative dobutamine stress echocardiography. J Am Soc Echocardiogr. 2014;27:880–7.

    Article  PubMed  Google Scholar 

  11. Bergenzaun L, Gudmundsson P, Ohlin H, During J, Ersson A, Ihrman L, et al. Assessing left ventricular systolic function in shock: evaluation of echocardiographic parameters in intensive care. Crit Care. 2011;15:R200.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Kopecna D, Briongos S, Castillo H, Moreno C, Recio M, Navas P, et al. Interobserver reliability of echocardiography for prognostication of normotensive patients with pulmonary embolism. Cardiovasc Ultrasound. 2014;12:29.

    Article  PubMed  PubMed Central  Google Scholar 

  13. De Geer L, Engvall J, Oscarsson A. Strain echocardiography in septic shock - a comparison with systolic and diastolic function parameters, cardiac biomarkers and outcome. Crit Care. 2015;19:857.

    Article  Google Scholar 

  14. Dellinger RP, Levy MM, Rhodes A, Annane D, Gerlach H, Opal SM, et al. Surviving sepsis campaign: international guidelines for management of severe sepsis and septic shock, 2012. Intensive Care Med. 2013;39:165–228.

    Article  CAS  PubMed  Google Scholar 

  15. Nagueh SF, Appleton CP, Gillebert TC, Marino PN, Oh JK, Smiseth OA, et al. Recommendations for the evaluation of left ventricular diastolic function by echocardiography. Eur J Echocardiogr. 2009;10:165–93.

    Article  PubMed  Google Scholar 

  16. Lang RM, Badano LP, Mor-Avi V, Afilalo J, Armstrong A, Ernande L, et al. Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the american society of echocardiography and the European association of cardiovascular imaging. J Am Soc Echocardiogr. 2015;28:1–39.

    Article  PubMed  Google Scholar 

  17. Dalen H, Thorstensen A, Aase SA, Ingul CB, Torp H, Vatten LJ, et al. Segmental and global longitudinal strain and strain rate based on echocardiography of 1266 healthy individuals: the HUNT study in Norway. Eur J Echocardiogr. 2010;11:176–83.

    Article  PubMed  Google Scholar 

  18. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.

    Article  CAS  PubMed  Google Scholar 

  19. Altman DG. Practical statistics for medical research. London, UK: Chapman & Hall; 1991.

    Google Scholar 

  20. Petrie MC, Hogg K, Caruana L, McMurray JJ. Poor concordance of commonly used echocardiographic measures of left ventricular diastolic function in patients with suspected heart failure but preserved systolic function: is there a reliable echocardiographic measure of diastolic dysfunction? Heart. 2004;90:511–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  21. Li C, Zhang J, Zhou C, Huang L, Tang H, Rao L. Will simultaneous measurement of E/e’ index facilitate the non-invasive assessment of left ventricular filling pressure in patients with non-valvular atrial fibrillation? Eur J Echocardiogr. 2010;11:296–301.

    Article  PubMed  Google Scholar 

  22. Yamamoto K, Nishimura RA, Chaliki HP, Appleton CP, Holmes Jr DR, Redfield MM. Determination of left ventricular filling pressure by Doppler echocardiography in patients with coronary artery disease: critical role of left ventricular systolic function. J Am Coll Cardiol. 1997;30:1819–26.

    Article  CAS  PubMed  Google Scholar 

  23. Sturgess DJ, Marwick TH, Joyce C, Jenkins C, Jones M, Masci P, et al. Prediction of hospital outcome in septic shock: a prospective comparison of tissue Doppler and cardiac biomarkers. Crit Care. 2010;14:R44.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Quintard H, Muller L, Philip I, Lena P, Ichai C. Influence of acute preload changes on mitral annulus velocity measured by tissue Doppler echocardiography in critically ill patients. J Clin Ultrasound. 2012;40:419–23.

    Article  PubMed  Google Scholar 

  25. Mahjoub Y, Benoit-Fallet H, Airapetian N, Lorne E, Levrard M, Seydi AA, et al. Improvement of left ventricular relaxation as assessed by tissue Doppler imaging in fluid-responsive critically ill septic patients. Intensive Care Med. 2012;38:1461–70.

    Article  PubMed  Google Scholar 

  26. Juhl-Olsen P, Hermansen JF, Frederiksen CA, Rasmussen LA, Jakobsen CJ, Sloth E. Positive end-expiratory pressure influences echocardiographic measures of diastolic function: a randomized, crossover study in cardiac surgery patients. Anesthesiology. 2013;119:1078–86.

    Article  PubMed  Google Scholar 

  27. St John Sutton M, Otterstat JE, Plappert T, Parker A, Sekarski D, Keane MG, et al. Quantitation of left ventricular volumes and ejection fraction in post-infarction patients from biplane and single plane two-dimensional echocardiograms: a prospective longitudinal study of 371 patients. Eur Heart J. 1998;19:808–16.

    Article  CAS  PubMed  Google Scholar 

  28. Sevilla Berrios RA, O’Horo JC, Velagapudi V, Pulido JN. Correlation of left ventricular systolic dysfunction determined by low ejection fraction and 30-day mortality in patients with severe sepsis and septic shock: a systematic review and meta-analysis. J Crit Care. 2014;29:495–9.

    Article  PubMed  Google Scholar 

  29. Maret E, Brudin L, Lindstrom L, Nylander E, Ohlsson JL, Engvall JE. Computer-assisted determination of left ventricular endocardial borders reduces variability in the echocardiographic assessment of ejection fraction. Cardiovasc Ultrasound. 2008;6:55.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Price S, Nicol E, Gibson DG, Evans TW. Echocardiography in the critically ill: current and potential roles. Intensive Care Med. 2006;32:48–59.

    Article  CAS  PubMed  Google Scholar 

  31. Hestenes SM, Halvorsen PS, Skulstad H, Remme EW, Espinoza A, Hyler S, et al. Advantages of strain echocardiography in assessment of myocardial function in severe sepsis: an experimental study. Crit Care Med. 2014;42:e432–40.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The study was supported by grants from the County Council of Östergötland, Sweden, and from the Heart-Lung Foundation, Sweden, grant number 20120449. The funding agent was not involved in study design, data collection, analysis or interpretation, or in writing or submitting the manuscript.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Lina De Geer.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

LDG participated in designing the study, was mainly responsible for collecting data, performed the echocardiographic measurements, participated in data analysis and interpretation, and was mainly responsible for writing the manuscript. AO participated in designing the study, in the collection, analysis and interpretation of data and in writing the manuscript. JE participated in designing the study, performed the echocardiograms and echocardiographic measurements, and participated in data interpretation and in writing the manuscript. All authors read and approved the final manuscript.

Additional file

Additional file 1:

Reproducibility of GLPS measurements in a four chamber view of the left ventricle. The given figures are based on an average from six segments of this particular view. From left to right: Measurements performed by observer 1, GLPS −19.8%; Measurements performed by observer 2, GLPS −18.2%; and measurements re-assessed by observer 2, GLPS −18.4%.

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

De Geer, L., Oscarsson, A. & Engvall, J. Variability in echocardiographic measurements of left ventricular function in septic shock patients. Cardiovasc Ultrasound 13, 19 (2015). https://doi.org/10.1186/s12947-015-0015-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12947-015-0015-6

Keywords